CVE-2015-3953
Summary
| CVE | CVE-2015-3953 |
|---|---|
| State | PUBLIC |
| Assigner | [email protected] |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2019-03-25 17:29:00 UTC |
| Updated | 2019-10-09 23:14:00 UTC |
| Description | Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue. |
Risk And Classification
Problem Types: CWE-798
NVD Known Affected Configurations (CPE 2.3)
| Type | Vendor | Product | Version | Update | Edition | Language |
|---|---|---|---|---|---|---|
| Hardware | Pifzer | Plum A 3 Infusion System | - | All | All | All |
| Operating System | Pifzer | Plum A 3 Infusion System Firmware | All | All | All | All |
| Hardware | Pifzer | Plum A Infusion System | - | All | All | All |
| Operating System | Pifzer | Plum A Infusion System Firmware | All | All | All | All |
| Hardware | Pifzer | Plum A 3 Infusion System | - | All | All | All |
| Hardware | Pifzer | Plum A 3 Infusion System | - | All | All | All |
| Operating System | Pifzer | Plum A 3 Infusion System Firmware | All | All | All | All |
| Hardware | Pifzer | Plum A Infusion System | - | All | All | All |
| Hardware | Pifzer | Plum A Infusion System | - | All | All | All |
| Operating System | Pifzer | Plum A Infusion System Firmware | All | All | All | All |
| Hardware | Pifzer | Symbiq Infusion System | - | All | All | All |
| Hardware | Pifzer | Symbiq Infusion System | - | All | All | All |
| Operating System | Pifzer | Symbiq Infusion System Firmware | All | All | All | All |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| Hospira Plum A+ and Symbiq Infusion Systems Vulnerabilities | ICS-CERT | MISC | ics-cert.us-cert.gov | Mitigation, Third Party Advisory, US Government Resource |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
No vendor comments have been submitted for this CVE.
There are currently no legacy QID mappings associated with this CVE.