CVE-2017-12718
Summary
| CVE | CVE-2017-12718 |
|---|---|
| State | PUBLIC |
| Assigner | [email protected] |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2018-02-15 10:29:00 UTC |
| Updated | 2019-10-09 23:23:00 UTC |
| Description | A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation. |
Risk And Classification
Problem Types: CWE-119
NVD Known Affected Configurations (CPE 2.3)
| Type | Vendor | Product | Version | Update | Edition | Language |
|---|---|---|---|---|---|---|
| Hardware | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | - | All | All | All |
| Hardware | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | - | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.1 | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.5 | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.6 | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.1 | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.5 | All | All | All |
| Operating System | Smiths-medical | Medfusion 4000 Wireless Syringe Infusion Pump | 1.6 | All | All | All |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| Smiths Medical Medfusion 4000 - 'DHCP' Denial of Service - Hardware dos Exploit | EXPLOIT-DB | www.exploit-db.com | Third Party Advisory, VDB Entry |
| Medfusion 4000 Wireless Syringe Infusion Pump ICSMA-17-250-02 Multiple Security Vulnerabilities | BID | www.securityfocus.com | Third Party Advisory, VDB Entry |
| Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A) | ICS-CERT | MISC | ics-cert.us-cert.gov | Third Party Advisory, US Government Resource |
| NXP Semiconductors MQX RTOS ICSA-17-285-04 Buffer Overflow and Denial Of Service Vulnerabilities | BID | www.securityfocus.com | Third Party Advisory, VDB Entry |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
No vendor comments have been submitted for this CVE.
There are currently no legacy QID mappings associated with this CVE.