CVE-2018-8868
Summary
| CVE | CVE-2018-8868 |
|---|---|
| State | PUBLIC |
| Assigner | [email protected] |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2018-07-03 01:29:00 UTC |
| Updated | 2019-10-09 23:42:00 UTC |
| Description | Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality. |
Risk And Classification
Problem Types: NVD-CWE-noinfo
NVD Known Affected Configurations (CPE 2.3)
| Type | Vendor | Product | Version | Update | Edition | Language |
|---|---|---|---|---|---|---|
| Hardware | Medtronic | 24950 Mycarelink Monitor | - | All | All | All |
| Hardware | Medtronic | 24950 Mycarelink Monitor | - | All | All | All |
| Operating System | Medtronic | 24950 Mycarelink Monitor Firmware | - | All | All | All |
| Operating System | Medtronic | 24950 Mycarelink Monitor Firmware | - | All | All | All |
| Hardware | Medtronic | 24952 Mycarelink Monitor | - | All | All | All |
| Hardware | Medtronic | 24952 Mycarelink Monitor | - | All | All | All |
| Operating System | Medtronic | 24952 Mycarelink Monitor Firmware | - | All | All | All |
| Operating System | Medtronic | 24952 Mycarelink Monitor Firmware | - | All | All | All |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| Medtronic MyCareLink Patient Monitor | CISA | MISC | ics-cert.us-cert.gov | Third Party Advisory, US Government Resource |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
No vendor comments have been submitted for this CVE.
There are currently no legacy QID mappings associated with this CVE.