DescriptionDräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

Risk And Classification

Primary CVSS: v4.0 7.2 HIGH from [email protected]

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Problem Types: CWE-798 | CWE-798 CWE-798 Use of Hard-coded Credentials


VersionSourceTypeScoreSeverityVector
4.0[email protected]Secondary7.2HIGHCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/C...
4.0CNACVSS7.2HIGHCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
3.1[email protected]Primary7.6HIGHCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
3.1CNACVSS7.6HIGHCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

CVSS v4.0 Breakdown

Attack Vector
Adjacent
Attack Complexity
Low
Attack Requirements
None
Privileges Required
None
User Interaction
None
Confidentiality
Low
Integrity
Low
Availability
High
Sub Conf.
None
Sub Integrity
None
Sub Availability
None

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

CVSS v3.1 Breakdown

Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
Low
Integrity
Low
Availability
High

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

Vendor Declared Affected Products

SourceVendorProductVersionPlatforms
CNA Drger SC 6002XL affected all versions VG2.3.1 custom Not specified
CNA Drger SC6802XL affected all versions custom Not specified
CNA Drger SC 7000 affected all versions custom Not specified
CNA Drger SC8000 affected all versions custom Not specified
CNA Drger SC90000 XL affected all versions custom Not specified

References

ReferenceSourceLinkTags
static.draeger.com/security [email protected] static.draeger.com
www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-hard-coded-credential... [email protected] www.vulncheck.com
CVE Program record CVE.ORG www.cve.org canonical
NVD vulnerability detail NVD nvd.nist.gov canonical, analysis

Vendor Comments And Credit

Discovery Credit

CNA: Jeroen Slobbe and Max Grim (en)

© CVE.report 2026 |

Use of this information constitutes acceptance for use in an AS IS condition. There are NO warranties, implied or otherwise, with regard to this information or its use. Any use of this information is at the user's risk. It is the responsibility of user to evaluate the accuracy, completeness or usefulness of any information, opinion, advice or other content. EACH USER WILL BE SOLELY RESPONSIBLE FOR ANY consequences of his or her direct or indirect use of this web site. ALL WARRANTIES OF ANY KIND ARE EXPRESSLY DISCLAIMED. This site will NOT BE LIABLE FOR ANY DIRECT, INDIRECT or any other kind of loss.