CVE-2020-15486
Summary
| CVE | CVE-2020-15486 |
|---|---|
| State | PUBLIC |
| Assigner | [email protected] |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2020-08-26 16:15:00 UTC |
| Updated | 2021-07-21 11:39:00 UTC |
| Description | An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity. |
Risk And Classification
Problem Types: NVD-CWE-noinfo
NVD Known Affected Configurations (CPE 2.3)
| Type | Vendor | Product | Version | Update | Edition | Language |
|---|---|---|---|---|---|---|
| Hardware | Drtrust | Electrocardiogram Pen | 2.00.08 | All | All | All |
| Hardware | Drtrust | Electrocardiogram Pen | 2.00.08 | All | All | All |
| Operating System | Drtrust | Electrocardiogram Pen Firmware | 2.00.08 | All | All | All |
| Operating System | Drtrust | Electrocardiogram Pen Firmware | 2.00.08 | All | All | All |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| Advisory | Payatu | MISC | payatu.com | Third Party Advisory |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
No vendor comments have been submitted for this CVE.
There are currently no legacy QID mappings associated with this CVE.