Mira Hormone Monitor, Mira Android App Missing authentication for critical function

Summary

CVECVE-2026-66098
StatePUBLISHED
Assignericscert
Source PriorityCVE Program / NVD first with legacy fallback
Published2026-08-11 22:18:49 UTC
Updated2026-08-12 13:17:23 UTC
DescriptionThe Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Risk And Classification

Primary CVSS: v4.0 7.1 HIGH from [email protected]

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Problem Types: CWE-306 | CWE-306 CWE-306


VersionSourceTypeScoreSeverityVector
4.0[email protected]Secondary7.1HIGHCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/C...
4.0CNACVSS7.1HIGHCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
3.1[email protected]Secondary6.5MEDIUMCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
3.1CNACVSS6.5MEDIUMCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

CVSS v4.0 Breakdown

Attack Vector
Adjacent
Attack Complexity
Low
Attack Requirements
None
Privileges Required
None
User Interaction
None
Confidentiality
None
Integrity
None
Availability
High
Sub Conf.
None
Sub Integrity
None
Sub Availability
None

CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

CVSS v3.1 Breakdown

Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
None
Availability
High

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Vendor Declared Affected Products

SourceVendorProductVersionPlatforms
CNA Quanovate Tech Inc. Operating As Mira Mira Care Mira Firmware affected 1.7.1.47 custom Not specified
CNA Quanovate Tech Inc. Operating As Mira Mira Care Mira Android App affected 4.5.15.4 custom Android

References

ReferenceSourceLinkTags
www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01 [email protected] www.cisa.gov
github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-2... [email protected] github.com
CVE Program record CVE.ORG www.cve.org canonical
NVD vulnerability detail NVD nvd.nist.gov canonical, analysis

Vendor Comments And Credit

Discovery Credit

CNA: Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech. (en)

Additional Advisory Data

Solutions

CNA: Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.

© CVE.report 2026

Use of this information constitutes acceptance for use in an AS IS condition. There are NO warranties, implied or otherwise, with regard to this information or its use. Any use of this information is at the user's risk. It is the responsibility of user to evaluate the accuracy, completeness or usefulness of any information, opinion, advice or other content. EACH USER WILL BE SOLELY RESPONSIBLE FOR ANY consequences of his or her direct or indirect use of this web site. ALL WARRANTIES OF ANY KIND ARE EXPRESSLY DISCLAIMED. This site will NOT BE LIABLE FOR ANY DIRECT, INDIRECT or any other kind of loss.