{"api_version":"1","generated_at":"2026-06-04T04:52:33+00:00","cve":"CVE-2019-25719","urls":{"html":"https://cve.report/CVE-2019-25719","api":"https://cve.report/api/cve/CVE-2019-25719.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2019-25719","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2019-25719"},"summary":{"title":"Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling","description":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.","state":"PUBLISHED","assigner":"VulnCheck","published_at":"2026-06-02 14:16:25","updated_at":"2026-06-03 16:16:17"},"problem_types":["CWE-924","CWE-924 CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel"],"metrics":[{"version":"4.0","source":"disclosure@vulncheck.com","type":"Secondary","score":"8.8","severity":"HIGH","vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.8,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"8.8","severity":"HIGH","vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"NETWORK","baseScore":8.8,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"disclosure@vulncheck.com","type":"Secondary","score":"8.6","severity":"HIGH","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","baseScore":8.6,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"8.6","severity":"HIGH","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","data":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":8.6,"baseSeverity":"HIGH","confidentialityImpact":"LOW","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","version":"3.1"}}],"references":[{"url":"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf","name":"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling","name":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://static.draeger.com/security","name":"https://static.draeger.com/security","refsource":"MITRE","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2019-25719","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25719","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"affected VG4.1.1 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"affected VG4.0.3 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"unaffected VG4.2 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","version":"affected VG4.1.1 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","version":"unaffected VG4.2 custom","platforms":[]}],"timeline":[],"solutions":[],"workarounds":[],"exploits":[],"credits":[],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":{"cve_year":"2019","cve_id":"25719","cve":"CVE-2019-25719","epss":"0.000170000","percentile":"0.044330000","score_date":"2026-06-03","updated_at":"2026-06-04 00:06:34"},"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"adp":[{"metrics":[{"other":{"content":{"id":"CVE-2019-25719","options":[{"Exploitation":"none"},{"Automatable":"yes"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","timestamp":"2026-06-02T14:56:18.300929Z","version":"2.0.3"},"type":"ssvc"}}],"providerMetadata":{"dateUpdated":"2026-06-02T14:56:35.870Z","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP"},"title":"CISA ADP Vulnrichment"}],"cna":{"affected":[{"defaultStatus":"unknown","product":"Infinity Acute Care System","vendor":"Dräger","versions":[{"lessThanOrEqual":"VG4.1.1","status":"affected","version":"0","versionType":"custom"},{"lessThanOrEqual":"VG4.0.3","status":"affected","version":"0","versionType":"custom"},{"status":"unaffected","version":"VG4.2","versionType":"custom"}]},{"defaultStatus":"unknown","product":"Standalone Infinity M540 patient monitor","vendor":"Dräger","versions":[{"lessThanOrEqual":"VG4.1.1","status":"affected","version":"0","versionType":"custom"},{"status":"unaffected","version":"VG4.2","versionType":"custom"}]}],"datePublic":"2019-09-12T00:00:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.</p>"}],"value":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"NETWORK","baseScore":8.8,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":8.6,"baseSeverity":"HIGH","confidentialityImpact":"LOW","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-924","description":"CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-06-03T15:43:03.688Z","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck"},"references":[{"tags":["vendor-advisory"],"url":"https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf"},{"tags":["third-party-advisory"],"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling"}],"source":{"discovery":"EXTERNAL"},"title":"Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling","x_generator":{"engine":"Vulnogram 1.0.2"}}},"cveMetadata":{"assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","assignerShortName":"VulnCheck","cveId":"CVE-2019-25719","datePublished":"2026-06-02T14:00:52.578Z","dateReserved":"2026-06-02T13:54:01.021Z","dateUpdated":"2026-06-03T15:43:03.688Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-06-02 14:16:25","lastModifiedDate":"2026-06-03 16:16:17","problem_types":["CWE-924","CWE-924 CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel"],"metrics":{"cvssMetricV40":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.8,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","baseScore":8.6,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH"},"exploitabilityScore":3.9,"impactScore":4.7}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2019","CveId":"25719","Ordinal":"1","Title":"Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Messag","CVE":"CVE-2019-25719","Year":"2019"},"notes":[{"CveYear":"2019","CveId":"25719","Ordinal":"1","NoteData":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.","Type":"Description","Title":"Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Messag"}]}}}