{"api_version":"1","generated_at":"2026-06-05T04:48:06+00:00","cve":"CVE-2019-25720","urls":{"html":"https://cve.report/CVE-2019-25720","api":"https://cve.report/api/cve/CVE-2019-25720.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2019-25720","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2019-25720"},"summary":{"title":"Dräger SC Monitoring Devices DoS via Malformed Network Packet","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.","state":"PUBLISHED","assigner":"VulnCheck","published_at":"2026-06-03 18:16:15","updated_at":"2026-06-04 15:29:14"},"problem_types":["CWE-1286","CWE-1286 CWE-1286 Improper Validation of Syntactic Correctness of Input"],"metrics":[{"version":"4.0","source":"disclosure@vulncheck.com","type":"Secondary","score":"7.1","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":7.1,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"7.1","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":7.1,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"disclosure@vulncheck.com","type":"Primary","score":"6.5","severity":"MEDIUM","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","baseScore":6.5,"baseSeverity":"MEDIUM","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"HIGH"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"6.5","severity":"MEDIUM","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","data":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"NONE","integrityImpact":"NONE","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","version":"3.1"}}],"references":[{"url":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf","name":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet","name":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2019-25720","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25720","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Dräger","product":"SC 6002XL","version":"affected SC 6002XL custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"SC6802XL","version":"affected SC6802XL custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"SC 7000","version":"affected SC 7000 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"SC8000","version":"affected SC8000 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"SC90000 XL","version":"affected SC90000 XL custom","platforms":[]}],"timeline":[],"solutions":[],"workarounds":[],"exploits":[],"credits":[{"source":"CNA","value":"Jeroen Slobbe and Max Grim","lang":"en"}],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":{"cve_year":"2019","cve_id":"25720","cve":"CVE-2019-25720","epss":"0.000240000","percentile":"0.072440000","score_date":"2026-06-04","updated_at":"2026-06-05 00:02:14"},"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"adp":[{"metrics":[{"other":{"content":{"id":"CVE-2019-25720","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","timestamp":"2026-06-03T17:44:08.218138Z","version":"2.0.3"},"type":"ssvc"}}],"providerMetadata":{"dateUpdated":"2026-06-03T17:44:47.456Z","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP"},"title":"CISA ADP Vulnrichment"}],"cna":{"affected":[{"defaultStatus":"unknown","product":"SC 6002XL","vendor":"Dräger","versions":[{"status":"affected","version":"SC 6002XL","versionType":"custom"}]},{"defaultStatus":"unknown","product":"SC6802XL","vendor":"Dräger","versions":[{"status":"affected","version":"SC6802XL","versionType":"custom"}]},{"defaultStatus":"unknown","product":"SC 7000","vendor":"Dräger","versions":[{"status":"affected","version":"SC 7000","versionType":"custom"}]},{"defaultStatus":"unknown","product":"SC8000","vendor":"Dräger","versions":[{"status":"affected","version":"SC8000","versionType":"custom"}]},{"defaultStatus":"unknown","product":"SC90000 XL","vendor":"Dräger","versions":[{"status":"affected","version":"SC90000 XL","versionType":"custom"}]}],"credits":[{"lang":"en","type":"finder","value":"Jeroen Slobbe and Max Grim"}],"datePublic":"2019-11-21T00:00:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.</p>"}],"value":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":7.1,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"NONE","integrityImpact":"NONE","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-1286","description":"CWE-1286 Improper Validation of Syntactic Correctness of Input","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-06-03T16:56:29.683Z","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck"},"references":[{"tags":["vendor-advisory"],"url":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf"},{"tags":["third-party-advisory"],"url":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet"}],"source":{"discovery":"EXTERNAL"},"title":"Dräger SC Monitoring Devices DoS via Malformed Network Packet"}},"cveMetadata":{"assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","assignerShortName":"VulnCheck","cveId":"CVE-2019-25720","datePublished":"2026-06-03T16:56:29.683Z","dateReserved":"2026-06-02T14:22:01.500Z","dateUpdated":"2026-06-03T17:44:47.456Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-06-03 18:16:15","lastModifiedDate":"2026-06-04 15:29:14","problem_types":["CWE-1286","CWE-1286 CWE-1286 Improper Validation of Syntactic Correctness of Input"],"metrics":{"cvssMetricV40":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":7.1,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"disclosure@vulncheck.com","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","baseScore":6.5,"baseSeverity":"MEDIUM","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"HIGH"},"exploitabilityScore":2.8,"impactScore":3.6}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2019","CveId":"25720","Ordinal":"1","Title":"Dräger SC Monitoring Devices DoS via Malformed Network Packet","CVE":"CVE-2019-25720","Year":"2019"},"notes":[{"CveYear":"2019","CveId":"25720","Ordinal":"1","NoteData":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.","Type":"Description","Title":"Dräger SC Monitoring Devices DoS via Malformed Network Packet"}]}}}