{"api_version":"1","generated_at":"2026-06-04T08:16:17+00:00","cve":"CVE-2019-25724","urls":{"html":"https://cve.report/CVE-2019-25724","api":"https://cve.report/api/cve/CVE-2019-25724.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2019-25724","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2019-25724"},"summary":{"title":"Dräger Infinity M300 VG2.x Network-Based Denial of Service","description":"Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.","state":"PUBLISHED","assigner":"VulnCheck","published_at":"2026-06-02 20:16:23","updated_at":"2026-06-02 20:16:23"},"problem_types":["CWE-400","CWE-400 CWE-400 Uncontrolled Resource Consumption"],"metrics":[{"version":"4.0","source":"disclosure@vulncheck.com","type":"Secondary","score":"7.1","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":7.1,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"7.1","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":7.1,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"disclosure@vulncheck.com","type":"Primary","score":"6.5","severity":"MEDIUM","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","baseScore":6.5,"baseSeverity":"MEDIUM","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"HIGH"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"6.5","severity":"MEDIUM","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","data":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"NONE","integrityImpact":"NONE","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","version":"3.1"}}],"references":[{"url":"https://static.draeger.com/security","name":"https://static.draeger.com/security","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service","name":"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2019-25724","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25724","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Dräger","product":"Infinity M300","version":"affected VG2.x and earlier VG2.3.1 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Infinity M300","version":"unaffected VG3.0 custom","platforms":[]}],"timeline":[],"solutions":[],"workarounds":[],"exploits":[],"credits":[],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":{"cve_year":"2019","cve_id":"25724","cve":"CVE-2019-25724","epss":"0.000160000","percentile":"0.035740000","score_date":"2026-06-03","updated_at":"2026-06-04 00:06:34"},"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"cna":{"affected":[{"defaultStatus":"unknown","product":"Infinity M300","vendor":"Dräger","versions":[{"lessThanOrEqual":"VG2.3.1","status":"affected","version":"VG2.x and earlier","versionType":"custom"},{"status":"unaffected","version":"VG3.0","versionType":"custom"}]}],"datePublic":"2020-07-24T00:00:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.</p>"}],"value":"Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":7.1,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":6.5,"baseSeverity":"MEDIUM","confidentialityImpact":"NONE","integrityImpact":"NONE","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-400","description":"CWE-400 Uncontrolled Resource Consumption","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-06-02T19:11:51.652Z","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck"},"references":[{"tags":["vendor-advisory"],"url":"https://static.draeger.com/security"},{"tags":["third-party-advisory"],"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service"}],"source":{"discovery":"INTERNAL"},"title":"Dräger Infinity M300 VG2.x Network-Based Denial of Service"}},"cveMetadata":{"assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","assignerShortName":"VulnCheck","cveId":"CVE-2019-25724","datePublished":"2026-06-02T19:11:51.652Z","dateReserved":"2026-06-02T19:05:51.385Z","dateUpdated":"2026-06-02T19:11:51.652Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-06-02 20:16:23","lastModifiedDate":"2026-06-02 20:16:23","problem_types":["CWE-400","CWE-400 CWE-400 Uncontrolled Resource Consumption"],"metrics":{"cvssMetricV40":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":7.1,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"NONE","vulnIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"disclosure@vulncheck.com","type":"Primary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H","baseScore":6.5,"baseSeverity":"MEDIUM","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"NONE","integrityImpact":"NONE","availabilityImpact":"HIGH"},"exploitabilityScore":2.8,"impactScore":3.6}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2019","CveId":"25724","Ordinal":"1","Title":"Dräger Infinity M300 VG2.x Network-Based Denial of Service","CVE":"CVE-2019-25724","Year":"2019"},"notes":[{"CveYear":"2019","CveId":"25724","Ordinal":"1","NoteData":"Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.","Type":"Description","Title":"Dräger Infinity M300 VG2.x Network-Based Denial of Service"}]}}}