{"api_version":"1","generated_at":"2026-06-03T21:13:57+00:00","cve":"CVE-2022-4992","urls":{"html":"https://cve.report/CVE-2022-4992","api":"https://cve.report/api/cve/CVE-2022-4992.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2022-4992","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2022-4992"},"summary":{"title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering","description":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.","state":"PUBLISHED","assigner":"VulnCheck","published_at":"2026-06-02 22:16:15","updated_at":"2026-06-03 20:16:17"},"problem_types":["CWE-345","CWE-345 CWE-345 Insufficient Verification of Data Authenticity"],"metrics":[{"version":"4.0","source":"disclosure@vulncheck.com","type":"Secondary","score":"8.8","severity":"HIGH","vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.8,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"8.8","severity":"HIGH","vector":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"NETWORK","baseScore":8.8,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"disclosure@vulncheck.com","type":"Secondary","score":"8.6","severity":"HIGH","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","baseScore":8.6,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"8.6","severity":"HIGH","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","data":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":8.6,"baseSeverity":"HIGH","confidentialityImpact":"LOW","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","version":"3.1"}}],"references":[{"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering","name":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf","name":"https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://static.draeger.com/security","name":"https://static.draeger.com/security","refsource":"MITRE","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2022-4992","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2022-4992","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"affected VG4.2 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"affected VG4.1.1 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Infinity Acute Care System","version":"affected VG4.0.3 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","version":"affected VG4.2 custom","platforms":[]},{"source":"CNA","vendor":"Dräger","product":"Standalone Infinity M540 patient monitor","version":"affected VG4.1.1 custom","platforms":[]}],"timeline":[],"solutions":[],"workarounds":[],"exploits":[],"credits":[],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":null,"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"adp":[{"metrics":[{"other":{"content":{"id":"CVE-2022-4992","options":[{"Exploitation":"none"},{"Automatable":"yes"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","timestamp":"2026-06-03T12:45:44.296578Z","version":"2.0.3"},"type":"ssvc"}}],"providerMetadata":{"dateUpdated":"2026-06-03T12:45:51.502Z","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP"},"title":"CISA ADP Vulnrichment"}],"cna":{"affected":[{"defaultStatus":"unknown","product":"Infinity Acute Care System","vendor":"Dräger","versions":[{"lessThan":"VG4.2","status":"affected","version":"0","versionType":"custom"},{"lessThan":"VG4.1.1","status":"affected","version":"0","versionType":"custom"},{"lessThan":"VG4.0.3","status":"affected","version":"0","versionType":"custom"}]},{"defaultStatus":"unknown","product":"Standalone Infinity M540 patient monitor","vendor":"Dräger","versions":[{"lessThan":"VG4.2","status":"affected","version":"0","versionType":"custom"},{"lessThan":"VG4.1.1","status":"affected","version":"0","versionType":"custom"}]}],"datePublic":"2022-01-24T00:00:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.</p>"}],"value":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"NETWORK","baseScore":8.8,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":8.6,"baseSeverity":"HIGH","confidentialityImpact":"LOW","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-345","description":"CWE-345 Insufficient Verification of Data Authenticity","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-06-03T19:56:58.455Z","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck"},"references":[{"tags":["vendor-advisory"],"url":"https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf"},{"tags":["third-party-advisory"],"url":"https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering"}],"source":{"discovery":"INTERNAL"},"title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering","x_generator":{"engine":"Vulnogram 1.0.2"}}},"cveMetadata":{"assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","assignerShortName":"VulnCheck","cveId":"CVE-2022-4992","datePublished":"2026-06-02T21:06:40.281Z","dateReserved":"2026-06-02T21:02:24.899Z","dateUpdated":"2026-06-03T19:56:58.455Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-06-02 22:16:15","lastModifiedDate":"2026-06-03 20:16:17","problem_types":["CWE-345","CWE-345 CWE-345 Insufficient Verification of Data Authenticity"],"metrics":{"cvssMetricV40":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.8,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","baseScore":8.6,"baseSeverity":"HIGH","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH"},"exploitabilityScore":3.9,"impactScore":4.7}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2022","CveId":"4992","Ordinal":"1","Title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling Do","CVE":"CVE-2022-4992","Year":"2022"},"notes":[{"CveYear":"2022","CveId":"4992","Ordinal":"1","NoteData":"Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.","Type":"Description","Title":"Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling Do"}]}}}