{"api_version":"1","generated_at":"2026-06-03T05:08:30+00:00","cve":"CVE-2026-0611","urls":{"html":"https://cve.report/CVE-2026-0611","api":"https://cve.report/api/cve/CVE-2026-0611.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2026-0611","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2026-0611"},"summary":{"title":"Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting","description":"Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.","state":"PUBLISHED","assigner":"VulnCheck","published_at":"2026-06-02 17:16:25","updated_at":"2026-06-02 20:16:31"},"problem_types":["CWE-306","CWE-306 Missing Authentication for Critical Function"],"metrics":[{"version":"4.0","source":"disclosure@vulncheck.com","type":"Secondary","score":"9.2","severity":"CRITICAL","vector":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":9.2,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"PRESENT","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"HIGH","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"9.2","severity":"CRITICAL","vector":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"PRESENT","attackVector":"NETWORK","baseScore":9.2,"baseSeverity":"CRITICAL","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"HIGH","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"disclosure@vulncheck.com","type":"Secondary","score":"9.8","severity":"CRITICAL","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","baseScore":9.8,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"9.8","severity":"CRITICAL","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","data":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":9.8,"baseSeverity":"CRITICAL","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","version":"3.1"}}],"references":[{"url":"https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel","name":"https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting","name":"https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting","refsource":"disclosure@vulncheck.com","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf","name":"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf","refsource":"134c704f-9b21-4f2e-91b3-4a467353bcc0","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2026-0611","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2026-0611","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Spacelabs Healthcare","product":"Sentinel","version":"affected 10.5.0 11.6.0 semver","platforms":[]}],"timeline":[],"solutions":[],"workarounds":[],"exploits":[],"credits":[{"source":"CNA","value":"Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp.","lang":"en"},{"source":"CNA","value":"Jan A. Rodriguez, Pentester, GM Sectec, Corp.","lang":"en"},{"source":"CNA","value":"VulnCheck","lang":"en"}],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":null,"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"adp":[{"metrics":[{"other":{"content":{"id":"CVE-2026-0611","options":[{"Exploitation":"poc"},{"Automatable":"yes"},{"Technical Impact":"total"}],"role":"CISA Coordinator","timestamp":"2026-06-02T19:13:38.769068Z","version":"2.0.3"},"type":"ssvc"}}],"providerMetadata":{"dateUpdated":"2026-06-02T19:13:46.749Z","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP"},"references":[{"tags":["exploit"],"url":"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"}],"title":"CISA ADP Vulnrichment"}],"cna":{"affected":[{"defaultStatus":"affected","product":"Sentinel","vendor":"Spacelabs Healthcare","versions":[{"lessThan":"11.6.0","status":"affected","version":"10.5.0","versionType":"semver"}]}],"credits":[{"lang":"en","type":"finder","value":"Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp."},{"lang":"en","type":"finder","value":"Jan A. Rodriguez, Pentester, GM Sectec, Corp."},{"lang":"en","type":"coordinator","value":"VulnCheck"}],"datePublic":"2026-06-02T00:00:00.000Z","descriptions":[{"lang":"en","value":"Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"PRESENT","attackVector":"NETWORK","baseScore":9.2,"baseSeverity":"CRITICAL","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"HIGH","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"HIGH","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"NETWORK","availabilityImpact":"HIGH","baseScore":9.8,"baseSeverity":"CRITICAL","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-306","description":"Missing Authentication for Critical Function","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-06-02T15:39:21.054Z","orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck"},"references":[{"tags":["vendor-advisory","patch"],"url":"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"},{"tags":["product"],"url":"https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel"},{"tags":["third-party-advisory"],"url":"https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting"}],"source":{"discovery":"EXTERNAL"},"title":"Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting","x_generator":{"engine":"vulncheck"}}},"cveMetadata":{"assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","assignerShortName":"VulnCheck","cveId":"CVE-2026-0611","datePublished":"2026-06-02T15:39:21.054Z","dateReserved":"2026-01-05T16:55:12.556Z","dateUpdated":"2026-06-02T19:13:46.749Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-06-02 17:16:25","lastModifiedDate":"2026-06-02 20:16:31","problem_types":["CWE-306","CWE-306 Missing Authentication for Critical Function"],"metrics":{"cvssMetricV40":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":9.2,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","attackRequirements":"PRESENT","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"HIGH","vulnAvailabilityImpact":"HIGH","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"disclosure@vulncheck.com","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","baseScore":9.8,"baseSeverity":"CRITICAL","attackVector":"NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"HIGH","integrityImpact":"HIGH","availabilityImpact":"HIGH"},"exploitabilityScore":3.9,"impactScore":5.9}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2026","CveId":"611","Ordinal":"1","Title":"Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RC","CVE":"CVE-2026-0611","Year":"2026"},"notes":[{"CveYear":"2026","CveId":"611","Ordinal":"1","NoteData":"Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.","Type":"Description","Title":"Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RC"}]}}}