{"api_version":"1","generated_at":"2026-08-12T18:06:04+00:00","cve":"CVE-2026-67558","urls":{"html":"https://cve.report/CVE-2026-67558","api":"https://cve.report/api/cve/CVE-2026-67558.json","docs":"https://cve.report/api","cve_org":"https://www.cve.org/CVERecord?id=CVE-2026-67558","nvd":"https://nvd.nist.gov/vuln/detail/CVE-2026-67558"},"summary":{"title":"Mira Hormone Monitor, Mira Android App Authentication bypass by spoofing","description":"The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view.","state":"PUBLISHED","assigner":"icscert","published_at":"2026-08-11 22:18:54","updated_at":"2026-08-12 16:17:14"},"problem_types":["CWE-290","CWE-290 CWE-290"],"metrics":[{"version":"4.0","source":"ics-cert@hq.dhs.gov","type":"Secondary","score":"8.2","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","data":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.2,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"ACTIVE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"NONE","subConfidentialityImpact":"HIGH","subIntegrityImpact":"LOW","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}},{"version":"4.0","source":"CNA","type":"CVSS","score":"8.2","severity":"HIGH","vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N","data":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":8.2,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"HIGH","subIntegrityImpact":"LOW","userInteraction":"ACTIVE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N","version":"4.0","vulnAvailabilityImpact":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"}},{"version":"3.1","source":"ics-cert@hq.dhs.gov","type":"Secondary","score":"7.4","severity":"HIGH","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","data":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","baseScore":7.4,"baseSeverity":"HIGH","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"REQUIRED","scope":"CHANGED","confidentialityImpact":"HIGH","integrityImpact":"LOW","availabilityImpact":"NONE"}},{"version":"3.1","source":"CNA","type":"CVSS","score":"7.4","severity":"HIGH","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","data":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"NONE","baseScore":7.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","version":"3.1"}}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01","name":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01","refsource":"ics-cert@hq.dhs.gov","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json","name":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json","refsource":"ics-cert@hq.dhs.gov","tags":[],"title":"","mime":"","httpstatus":"","archivestatus":"0"},{"url":"https://www.cve.org/CVERecord?id=CVE-2026-67558","name":"CVE Program record","refsource":"CVE.ORG","tags":["canonical"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2026-67558","name":"NVD vulnerability detail","refsource":"NVD","tags":["canonical","analysis"]}],"affected":[{"source":"CNA","vendor":"Quanovate Tech Inc. (operating as Mira / Mira Care)","product":"Mira Firmware","version":"affected 1.7.1.47 custom","platforms":[]},{"source":"CNA","vendor":"Quanovate Tech Inc. (operating as Mira / Mira Care)","product":"Mira Android App","version":"affected 4.5.15.4 custom","platforms":["Android"]}],"timeline":[],"solutions":[{"source":"CNA","title":"","value":"Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.","time":"","lang":"en"}],"workarounds":[],"exploits":[],"credits":[{"source":"CNA","value":"Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech.","lang":"en"}],"nvd_cpes":[],"vendor_comments":[],"enrichments":{"kev":null,"epss":null,"legacy_qids":[]},"source_records":{"cve_program":{"containers":{"adp":[{"metrics":[{"other":{"content":{"id":"CVE-2026-67558","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","timestamp":"2026-08-12T15:19:26.583620Z","version":"2.0.3"},"type":"ssvc"}}],"providerMetadata":{"dateUpdated":"2026-08-12T15:19:42.642Z","orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP"},"title":"CISA ADP Vulnrichment"}],"cna":{"affected":[{"defaultStatus":"unaffected","product":"Mira Firmware","vendor":"Quanovate Tech Inc. (operating as Mira / Mira Care)","versions":[{"status":"affected","version":"1.7.1.47","versionType":"custom"}]},{"defaultStatus":"unaffected","platforms":["Android"],"product":"Mira Android App","vendor":"Quanovate Tech Inc. (operating as Mira / Mira Care)","versions":[{"status":"affected","version":"4.5.15.4","versionType":"custom"}]}],"credits":[{"lang":"en","type":"finder","value":"Gigi Xiaoqing Liu, Muzzammil Mohammed, Narmina Karimova, and En Mong of Northeastern University SPQR Lab reported this vulnerability to Quanovate Tech."}],"datePublic":"2026-08-11T17:00:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view."}],"value":"The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"NONE","attackVector":"ADJACENT","baseScore":8.2,"baseSeverity":"HIGH","exploitMaturity":"NOT_DEFINED","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"HIGH","subIntegrityImpact":"LOW","userInteraction":"ACTIVE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N","version":"4.0","vulnAvailabilityImpact":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"LOW","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"NONE","baseScore":7.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"LOW","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-290","description":"CWE-290","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2026-08-11T21:27:03.214Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01"},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json"}],"solutions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."}],"value":"Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required."}],"source":{"advisory":"ICSMA-26-223-01","discovery":"EXTERNAL"},"title":"Mira Hormone Monitor, Mira Android App Authentication bypass by spoofing","x_generator":{"engine":"Vulnogram 1.0.4"}}},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2026-67558","datePublished":"2026-08-11T21:27:03.214Z","dateReserved":"2026-08-03T16:54:56.472Z","dateUpdated":"2026-08-12T15:19:42.642Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.2"},"nvd":{"publishedDate":"2026-08-11 22:18:54","lastModifiedDate":"2026-08-12 16:17:14","problem_types":["CWE-290","CWE-290 CWE-290"],"metrics":{"cvssMetricV40":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","cvssData":{"version":"4.0","vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","baseScore":8.2,"baseSeverity":"HIGH","attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"ACTIVE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"LOW","vulnAvailabilityImpact":"NONE","subConfidentialityImpact":"HIGH","subIntegrityImpact":"LOW","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","confidentialityRequirement":"NOT_DEFINED","integrityRequirement":"NOT_DEFINED","availabilityRequirement":"NOT_DEFINED","modifiedAttackVector":"NOT_DEFINED","modifiedAttackComplexity":"NOT_DEFINED","modifiedAttackRequirements":"NOT_DEFINED","modifiedPrivilegesRequired":"NOT_DEFINED","modifiedUserInteraction":"NOT_DEFINED","modifiedVulnConfidentialityImpact":"NOT_DEFINED","modifiedVulnIntegrityImpact":"NOT_DEFINED","modifiedVulnAvailabilityImpact":"NOT_DEFINED","modifiedSubConfidentialityImpact":"NOT_DEFINED","modifiedSubIntegrityImpact":"NOT_DEFINED","modifiedSubAvailabilityImpact":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED"}}],"cvssMetricV31":[{"source":"ics-cert@hq.dhs.gov","type":"Secondary","cvssData":{"version":"3.1","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N","baseScore":7.4,"baseSeverity":"HIGH","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"REQUIRED","scope":"CHANGED","confidentialityImpact":"HIGH","integrityImpact":"LOW","availabilityImpact":"NONE"},"exploitabilityScore":2.1,"impactScore":4.7}],"ssvcV203":[{"source":"134c704f-9b21-4f2e-91b3-4a467353bcc0","ssvcData":{"timestamp":"2026-08-12T15:19:26.583620Z","id":"CVE-2026-67558","options":[{"exploitation":"none"},{"automatable":"no"},{"technicalImpact":"partial"}],"role":"CISA Coordinator","version":"2.0.3"}}]},"configurations":[]},"legacy_mitre":{"record":{"CveYear":"2026","CveId":"67558","Ordinal":"1","Title":"Mira Hormone Monitor, Mira Android App Authentication bypass by ","CVE":"CVE-2026-67558","Year":"2026"},"notes":[{"CveYear":"2026","CveId":"67558","Ordinal":"1","NoteData":"The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view.","Type":"Description","Title":"Mira Hormone Monitor, Mira Android App Authentication bypass by "}]}}}