QID 590903
QID 590903: Illumina Local Run Manager Multiple Vulnerabilities (ICSA-22-153-02)
AFFECTED PRODUCTS
The following devices and instruments using LRM software are affected:
Illumina In Vitro Diagnostic (IVD) devices:
NextSeq 550Dx: LRM Versions 1.3 to 3.1
MiSeq Dx: LRM Versions 1.3 to 3.1
Researcher Use Only (ROU) instruments:
NextSeq 500 Instrument: LRM Versions 1.3 to 3.1
NextSeq 550 Instrument: LRM Versions 1.3 to 3.1
MiSeq Instrument: LRM Versions 1.3 to 3.1
iSeq 100 Instrument: LRM Versions 1.3 to 3.1
MiniSeq Instrument: LRM Versions 1.3 to 3.1
QID Detection Logic (Authenticated)
QID checks for the Vulnerable version using windows registry keys
Successful exploitation of these vulnerabilities may allow an unauthenticated malicious actor to take control of the affected product remotely and take any action at the operating system level. An attacker could impact settings, configurations, software, or data on the affected product and interact through the affected product with the connected network..
Customers are advised to refer to CERT MITIGATIONS section ICSA-22-153-02 for affected packages and patching details.
- ICSA-22-153-02 -
www.us-cert.gov/ics/advisories/ICSA-22-153-02