Dräger Infinity M300 VG2.x Network-Based Denial of Service
Summary
| CVE | CVE-2019-25724 |
|---|---|
| State | PUBLISHED |
| Assigner | VulnCheck |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2026-06-02 20:16:23 UTC |
| Updated | 2026-06-02 20:16:23 UTC |
| Description | Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered. |
Risk And Classification
Primary CVSS: v4.0 7.1 HIGH from [email protected]
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
EPSS: 0.000160000 probability, percentile 0.035740000 (date 2026-06-03)
Problem Types: CWE-400 | CWE-400 CWE-400 Uncontrolled Resource Consumption
| Version | Source | Type | Score | Severity | Vector |
|---|---|---|---|---|---|
| 4.0 | [email protected] | Secondary | 7.1 | HIGH | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/C... |
| 4.0 | CNA | CVSS | 7.1 | HIGH | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N |
| 3.1 | [email protected] | Primary | 6.5 | MEDIUM | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| 3.1 | CNA | CVSS | 6.5 | MEDIUM | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
CVSS v4.0 Breakdown
Attack Vector
AdjacentAttack Complexity
LowAttack Requirements
NonePrivileges Required
NoneUser Interaction
NoneConfidentiality
NoneIntegrity
NoneAvailability
HighSub Conf.
NoneSub Integrity
NoneSub Availability
NoneCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS v3.1 Breakdown
Attack Vector
AdjacentAttack Complexity
LowPrivileges Required
NoneUser Interaction
NoneScope
UnchangedConfidentiality
NoneIntegrity
NoneAvailability
HighCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Vendor Declared Affected Products
| Source | Vendor | Product | Version | Platforms |
|---|---|---|---|---|
| CNA | Drger | Infinity M300 | affected VG2.x and earlier VG2.3.1 custom | Not specified |
| CNA | Drger | Infinity M300 | unaffected VG3.0 custom | Not specified |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| static.draeger.com/security | [email protected] | static.draeger.com | |
| www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of... | [email protected] | www.vulncheck.com | |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
No vendor comments have been submitted for this CVE.
There are currently no legacy QID mappings associated with this CVE.