Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Summary
| CVE | CVE-2026-0611 |
|---|---|
| State | PUBLISHED |
| Assigner | VulnCheck |
| Source Priority | CVE Program / NVD first with legacy fallback |
| Published | 2026-06-02 17:16:25 UTC |
| Updated | 2026-06-02 20:16:31 UTC |
| Description | Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes. |
Risk And Classification
Primary CVSS: v4.0 9.2 CRITICAL from [email protected]
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Problem Types: CWE-306 | CWE-306 Missing Authentication for Critical Function
| Version | Source | Type | Score | Severity | Vector |
|---|---|---|---|---|---|
| 4.0 | [email protected] | Secondary | 9.2 | CRITICAL | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/C... |
| 4.0 | CNA | CVSS | 9.2 | CRITICAL | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| 3.1 | [email protected] | Secondary | 9.8 | CRITICAL | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| 3.1 | CNA | CVSS | 9.8 | CRITICAL | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
CVSS v4.0 Breakdown
Attack Vector
NetworkAttack Complexity
LowAttack Requirements
PresentPrivileges Required
NoneUser Interaction
NoneConfidentiality
HighIntegrity
HighAvailability
HighSub Conf.
NoneSub Integrity
NoneSub Availability
NoneCVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS v3.1 Breakdown
Attack Vector
NetworkAttack Complexity
LowPrivileges Required
NoneUser Interaction
NoneScope
UnchangedConfidentiality
HighIntegrity
HighAvailability
HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vendor Declared Affected Products
| Source | Vendor | Product | Version | Platforms |
|---|---|---|---|---|
| CNA | Spacelabs Healthcare | Sentinel | affected 10.5.0 11.6.0 semver | Not specified |
References
| Reference | Source | Link | Tags |
|---|---|---|---|
| spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel | [email protected] | spacelabshealthcare.com | |
| www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rc... | [email protected] | www.vulncheck.com | |
| spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory... | 134c704f-9b21-4f2e-91b3-4a467353bcc0 | spacelabshealthcare.com | |
| CVE Program record | CVE.ORG | www.cve.org | canonical |
| NVD vulnerability detail | NVD | nvd.nist.gov | canonical, analysis |
Vendor Comments And Credit
Discovery Credit
CNA: Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp. (en)
CNA: Jan A. Rodriguez, Pentester, GM Sectec, Corp. (en)
CNA: VulnCheck (en)
There are currently no legacy QID mappings associated with this CVE.